Low Flow Coriolis Mass Flow Meters / Controllers for Liquid and Gases
The mini CORI-FLOW? Mass Flow Meters and Controllers from Bronkhorst? are precise and compact instruments based on the Coriolis measurement principle. They are designed to meet the needs of the low-flow market.
The Bronkhorst? model M15 Mass Flow Meter (MFM) is suitable for highly accurate measurement of gas or liquid flow in the range of 0 to 300 kg/h, which corresponds to 0 to 4000 ln/min when used with nitrogen. These instruments can operate at pressures up to 100 bar, with the possibility of higher pressure ratings upon request.
The instruments are equipped with a robust IP65 weatherproof housing and are optionally available with ATEX approval for use in Zone 2 hazardous areas.
The instrument contains a microprocessor-based PC-board with signal and fieldbus conversion, as well as a PID controller for optional mass flow control. This control can be achieved by means of a separately mounted control valve or pump.
Technical specifications
Measurement / control system | |
---|---|
Flow rates | Liquid: 0…300 kg/h (nominal flow rate: 100 kg/h); Gas: 0…4000 ln/min (N2); Full Scale (FS) value is user-configurable |
Mass flow accuracy | Liquid: ±0,2% Rd; Gas: ±0,5% Rd |
Repeatability | ±?0,05?%?of rate ± ?(ZS* x 100/actual flow)% |
Turndown ratio | up to 1:100 |
Zero stability (ZS) | < ±?50?g/h (Guaranteed at constant temperature and for unchanging process and environment conditions.) |
Response time (sensor) | ≤?200?msec |
Temperature effect | on zero: < 5 g/h/°C; on span: < 0,001% Rd/°C; self heating (at zero flow): < 10°C (Depends on flow rate, heat capacity fluid, T amb., T fluid and cooling capacity.) |
Operating temperature | 0?… 70?°C?; for ATEX Cat.3, Zone 2 max. 50°C |
Mounting | any position, attitude sensitivity negligible. External shocks or vibrations should be avoided. |
Temperature accuracy | ±?0,5?°C |
Density accuracy | < ±?5?kg/m??(at full scale flow) |
Leak integrity, outboard | tested < 2 x 10-9?mbar l/s He |
Warm-up time | >?30 min?for optimum accuracy |
Mechanical parts | |
---|---|
Sensor | single tube, DN 3.1 |
Material (wetted parts) | stainless steel 316L / 1.4404; |
Housing | stainless steel 430F |
Pressure rating (PN) | 100?bar abs?, higher on request |
Process connections | compression type or face seal (VCR/VCO) couplings, or Tri-Clamp flanges (welded) |
Seals | metal |
Weight | 4,7?kg |
Ingress protection | IP65 (weatherproof) |
Electrical properties | |
---|---|
Power supply | +15…24 Vdc +/- 10% Max. ripple recommended: 50 mV tt |
Max. power consumption | meter: max. 3 W; controller: max 7 W |
Analog output | 0…5 (10) Vdc, min. load impedance > 2 k?; 0 (4)…20 mA (sourcing), max. load impedance < 375 ? |
Analog setpoint | 0…5 (10) Vdc, min. load impedance > 100 k? 0 (4)…20 mA (sourcing), max. load impedance ~ 250 ? |
Digital communication | standard: RS232; options: PROFIBUS DP, DeviceNet?, Modbus RTU or FLOW-BUS |
Electrical connection | |
---|---|
Analog/RS232 | 8-pin DIN (male); |
PROFIBUS DP | bus: 5-pin M12 (female); power: 8-pin DIN (male); |
CANopen? / DeviceNet? | 5-pin M12 (male); |
FLOW-BUS/Modbus-RTU/ASCII | 5-pin M12 (male); |
Control valve options | |
---|---|
M15+C0I: Gas flow control valve | Kv-max= 6,6 x 10-2 |
M15+C2I: Liquid flow control valve | Kv-max= 2,3 x 10-3 |
M15+C5I: Gas/Liquid flow control valve | Kv-max= 6,6 x 10-2 |
M15+F-004AI: Gas/Liquid flow control | Kv-max= 3,0 x 10-1 |
M15+F-004BI: Gas/Liquid flow control | Kv-max= 1,0 |
Technical specifications subject to change without notice.